Safety of mechanical chest compression devices in cardiac arrest: a randomised virtopsy study with the AutoPulse®

ISRCTN ISRCTN75393297
DOI https://doi.org/10.1186/ISRCTN75393297
Secondary identifying numbers N/A
Submission date
10/06/2009
Registration date
11/09/2009
Last edited
21/01/2019
Recruitment status
No longer recruiting
Overall study status
Completed
Condition category
Circulatory System
Prospectively registered
Protocol
Statistical analysis plan
Results
Individual participant data

Plain English Summary

Background and study aims
Cardio-pulmonary resuscitation (CPR) is a first aid technique that can be used if someone’s heart has stopped (cardiac arrest). Chest compression during CPR is hard work and many rescuers fail to achieve the required compression depth and compression rate for longer than 1-2 minutes. Also, compressions are interrupted both for good or bad reasons, possibly with a negative impact on outcome. It is not surprising that mechanical chest compression devices have been developed which could give chest compressions without interruption and without fatigue. There were great expectations that these devices could improve survival from cardiac arrest. Unfortunately, at the time of design of this study, no studies had showed that expected benefit, rather the opposite. Since our current study started, several studies have been published that also failed to show a clear benefit of mechanical chest compression compared to manual compressions.
One possible explanation could be that the potential benefit was counteracted by possibly too forceful compressions or by possible incorrect positioning of the device, causing damage to internal organs. Such damage was not demonstrated convincingly, only in some anecdotal cases. The purpose of our study is to specifically look for damage to bone structures and internal organs by means of postmortem CT scans, autopsy, or evaluation of the clinical course of the patient after successful resuscitation. All analyses are done by observers that are ‘blind’, i.e. not aware of the method of chest compression that was used in a particular patient.
There are several different ways to administer chest compressions and to achieve the desired propelling of blood. One of the ways is a device that rhythmically applies force to a band over the chest, causing a more general compression of the whole chest, rather than only pushing the sternum (breastbone) inward. This device, Autopulse, is subject of this study and is compared to manual chest compression by in-hospital rescuers.

Who can participate?
Patients aged 18 years or over, either sex, are included in the study when in cardiac arrest and requiring full CPR, which is ongoing when the resuscitation team arrives with the study device.

What does the study involve?
When the need for CPR arises in a patient, chest compressions are performed with either the Autopulse or with conventional manual chest compressions. The method used is chosen at random. If the patient does not survive, the family is approached for permission to perform a postmortem CT scan (which does not open the body but only involves an X-ray study of the intact body) or an autopsy. As the patient is unconscious from the outset of cardiac arrest, the patient is not aware of the use of either devices.

What are the possible benefits and risks of participating?
The benefit could be that more patients survive and no excess damage is done. The risk could be that the excess damage is demonstrated. As these devices are commercially available and applied widely, understanding the possible risks of its use is of great importance.

Where is the study run from?
Academic Medical Centre (AMC) (Netherlands).

When is the study starting and how long is it expected to run for?
The study ran from June 2009 to May 2014.

Who is funding the study?
Zoll Medical Corporation (USA) and AMC Medical Research BV (Netherlands).

Who is the main contact?
Dr Rudolph Koster

Contact information

Dr Rudolph Koster
Scientific

Academic Medical Centre (AMC)
Department of Cardiology
Meibergdreef 9
Amsterdam
1105 AZ
Netherlands

Study information

Study designSingle-centre randomised controlled part-blinded non-inferiority clinical trial
Primary study designInterventional
Secondary study designRandomised controlled trial
Study setting(s)Hospital
Study typeTreatment
Participant information sheet Not available in web format, please use the contact details below to request a patient information sheet
Scientific titleSafety of use of the AutoPulse®: a mechanical chest compression device in in-hospital cardiac arrest: a randomised virtopsy study to investigate the possible damage from chest compression administered by manual force compared with the AutoPulse® assessed blind by (post-mortem) computed tomography (CT) scanning
Study acronymArrest 16A
Study hypothesisUse of the AutoPulse® device during cardio-pulmonary resuscitation compared with manual chest compression, is not associated with a significant increased probability of severe damage: immediately life-endangering or severely debilitating damage to parenchymal organs (liver, spleen, lungs), gastro-intestinal rupture, tension pneumothorax, myocardial rupture or aortic rupture, spinal, vertebral damage or flail chest.

On 28/05/2014 the anticipated end date was changed from 01/06/2011 to 26/05/2014.
Ethics approval(s)Medical Ethics Board of the Academic Medical Center, Amsterdam, 08/11/2007, ref: 07-249
ConditionCirculatory arrest
InterventionInvestigational treatment: chest compressions with AutoPulse® device
Control treatment: manual chest compressions with audio-visual feedback from displacement sternum transducer

The possible damage will be assessed by blinded investigators from the post-mortem CT scan, the autopsy if available or the clinical course of (initial) survivors, if no clinically indicated CT scan is available. The seriousness of the complications observed will be distinguished in three levels of severity: life-threatening, consequential or insignificant. The damage will also be classified as related to cause of arrest or CPR related or uncertain.

Patients will be followed-up until discharge or death in-hospital.
Intervention typeDevice
Pharmaceutical study type(s)
PhaseNot Applicable
Drug / device / biological / vaccine name(s)
Primary outcome measureCPR-related severe damage, defined as severe, life-threatening damage to a visceral organ in the chest or abdomen, but also including large vessels and vertebrae. Damage will be classified as primary endpoint by the data and safety committee, based on the available information, including (post-mortem) CT scans, autopsy, if performed, and clinical information in (initial) survivors.

Assessment will be done after in-hospital death or at hospital discharge.
Secondary outcome measuresDamage to bony structures of the chest wall (sternum and ribs). Assessment will be done after in-hospital death or at hospital discharge.
Overall study start date12/06/2009
Overall study end date26/05/2014

Eligibility

Participant type(s)Patient
Age groupAdult
Lower age limit18 Years
SexBoth
Target number of participantsNon-inferiority design: 2 x 112 patients needed
Participant inclusion criteriaPatients (aged 18 years or over, either sex) are included in the study when in circulatory arrest, requiring full cardio-pulmonary resuscitation, which is ongoing when the resuscitation team arrives with the study device.
Participant exclusion criteria1. Patients under 18 years of age
2. Trauma is the cause of circulatory arrest
Recruitment start date12/06/2009
Recruitment end date26/05/2014

Locations

Countries of recruitment

  • Netherlands

Study participating centre

Academic Medical Centre (AMC)
Amsterdam
1105 AZ
Netherlands

Sponsor information

Academic Medical Centre (AMC) (Netherlands)
Hospital/treatment centre

c/o R.W. Koster
Department of Cardiology
Meibergdreef 9
Amsterdam
1105 AZ
Netherlands

Website http://www.amc.nl/?pid=2581
ROR logo "ROR" https://ror.org/03t4gr691

Funders

Funder type

Industry

Zoll Medical Corporation (USA) - restricted grant

No information available

AMC Medical Research BV (Netherlands)

No information available

Results and Publications

Intention to publish date
Individual participant data (IPD) Intention to shareNo
IPD sharing plan summaryNot provided at time of registration
Publication and dissemination planNot provided at time of registration
IPD sharing plan

Study outputs

Output type Details Date created Date added Peer reviewed? Patient-facing?
Results article results 21/10/2017 21/01/2019 Yes No